KLD Mono&Multiple Water Quality Analyzer
On-line Analyzer for Ammonia Monitoring (Nessler’s Reagent Method)

On-line Analyzer for Ammonia Monitoring (Nessler’s Reagent Method)

    KLD-C-NH3-N is an automatic on-line analyzer independently developed by KLD. Using imported CPU, highly precise signal processor and simple but steady loop flow system to effectively reduce signal fluctuation and to ensure the accuracy and stability of data. Complete separation between electronics and hydraulics allows long-term & stable running. It is suitable for several types of water matrix like river water, surface water and industrial effluent.

Hits:346
Product Presentation

    KLD-C-NH3-N is an automatic on-line analyzer independently developed by KLD. Using imported CPU, highly precise signal processor and simple but steady loop flow system to effectively reduce signal fluctuation and to ensure the accuracy and stability of data. Complete separation between electronics and hydraulics allows long-term & stable running. It is suitable for several types of water matrix like river water, surface water and industrial effluent.


Features

1.      Robust & Reliable

Designed for on-line industrial and environmental applications. Complete separation between electronics and hydraulics allows long-term stable operations.

2.      Precise Reagent Dosage

CPU controlled photoelectric dosing system for accurate reagent dosage. Simple & steady structure ensures long-term service life.

3.      Easy Installation

Delivered after a long & successful series of factory tests, it is ready for installation and setup. To start monitoring just need to connect reagent tubes, sample tube, water tube, waste tube and power supply, then set up the parameters.

4.      Automatic Calibration

The analyzer performs calibration according to the calibration interval and type set up by users. If the result exceeds selected limits, the original calibration parameters will not be updated and an alarm signal will be outputted.

5.      Automatic Monitoring

Measuring interval can be freely set up by users. Between two measurements the analyzer remains in standby.


Benefits

?  Automatic monitoring system gives alarm signal in abnormal condition. Easy and low maintenance; Low operating cost.

?  4-20mA analog signal output; standard RS232, RS485 digital signal output, MODBUS protocol.

?  New signal processor reduces signal fluctuation to ensure the accuracy and stability of data.

?  Highly precise peristaltic pump extends the service life of pump tubes.

?  Automatic restart when normal power supply after accidental power down.

?  Remote control with local or remote PC connection.

?  Low reagent consumption; short preparation time.

?  Automatic diagnosis function. Identification and alarm for reagent shortage.

?  Instrument operating signal output function.

?  3000 groups of data storage and USB download for analysis data and running log.

?  Easy operation, no special training is required.

?  Controllable running of water pump and free setup for the sampling time of the pump.

?  Security management function, automatic record and storage for all control operations. Read-only storage.

?  2-level operation & management authorization: The system administrator can have all system operation and general operators can only perform routine maintenance & operation.


Measuring Principle

The sample after proper filtration is pumped into the reactor, where R1(potassium sodium tartrate, disodium hydrogen phosphate and sodium hydroxide) and R2 (sodium hydroxide, potassium iodide and mercury iodide) are injected. After fully mixing and reaction, the analyzer measures ABS value of light reddish brown complexes from the reaction at the specific wavelength, then calculates sample concentration according to the stored calibration data.


Specifications

Measuring Principle

Colorimetric, Nessler’s Reagent Method

Operating Management

Security management and 2-level operation management function

Measuring Type

Manual or automatic analysis with interval

Measuring Interval

Programmable; reverse control available by RS232

Measuring Time

12 minutes

Measuring Range

0.01~5/10/20/50/100mg/L, other ranges available on request

Calibration Type

Manual or automatic analysis with interval

Precision

≤ 5%

Accuracy

±5%

Zero Drift

5%F.S

Range Drift

5%F.S

Signal Output

Standard 4-20mA analog output,

Standard RS232/RS485 digital output, MODBUS Protocol

Alarm

High limit alarm, low limit alarm, reagent shortage alarm,   calibration error alarm (standard)

Sample and Waste Delivery

Pressure free

Reagent Temperature

10-30°C

Reagent Replacement

3~4 weeks (depending on the ambient temperature)

Ambient temperature

5-40°C

Protection

IP55

MTBF

720h

Display

Color touch screen (language: Chinese Simplified, Chinese   Traditional, English)

Power supply

220V AC±10%; power 90W

Weight

60kg (without reagents)

Dimension

1600×500×400mm(H×W×D) (with the height of base)


Optional Functions

Dual-channel

One analyzer to monitor on-line for 2 outlets.

Reagent Cooling

Longer reagent service life and less reagent replacing.

Alarm Function

DI water low limit alarm and waste liquid high limit alarm function.

Automatic Dilution Function

Automatic dilution function for high concentration sample.

Mini Printer

Print data after measurement.


    Tags:On-line Analyzer for Ammonia Monitoring (Nessler’s Reagent Method)